Frequency of Pipette Preventative Maintenance and Calibration

Most major manufacturers recommend pipette preventative maintenance (PM) and calibration service at least every 12 months, based on use. Twelve months is a good starting interval and should include:

  • A liquid leak test, 12-point inspection
  • Cleaning
  • Manufacturer recommended pipette preventative maintenance
  • Replacement of common PM parts, such as the seal
  • lubrication of the piston assembly with manufacturer lubricant (as required by the manufacturer)
  • And calibration adjustments, as needed. Calibration should be a last resort after all other components are deemed good.

Printable Calibration Log
Printable Calibration Stickers (for use with Avery 5960 labels)

pipette disassembly for preventative maintenance

Disclaimer: Any action you take using the information on this website is strictly at your own risk. The information herein does not constitute professional advice and is general in nature. We make no warranty that this information will meet your requirements, be safe, accurate, or error-free. Pipette Supplies, Inc. is not responsible for any errors or omissions, any results obtained from the use of this information, or any loss or damage arising out of the use of this information. This site is for educational purposes only.