These adjustment instructions cover the following Nichiryo models: Notes: Tool Needed: – Nichiryo/Oxford Hex Head Wrench Calibration Tool; Item#: NI 5G2-HEX Step #1: Hold down the ejector button of the pipette and turn the locking lever/collar to expose the adjustment screw channel (figure #0). Step #2: Adjust the pipette to 75-80% of the nominal volume